Brooks-instrument 0254 User Manual Page 87

  • Download
  • Add to my manuals
  • Print
  • Page
    / 124
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 86
C-9
Model 0254
Section C Serial Communications Protocol
Installation and Operation Manual
X-SE-0254-eng
Part Number: 541B129AAG
September, 2010
C-4-1-5 Message Serial Character Pacing Controls
These commands provide for the terminal or host to suspend character
sending or to re-enable the sending to continue. This facility is particularly
useful when the instrument sends large data amounts, such as logged
information.
Serial Character Pacing
This command acts as an XOF to temporarily suspend unit from sending
further characters:
HOST SEND
HOST SEND
AZH<cr> Non-network
AZ[yyyyy]H<cr> Network
RESPONSE - none
This command acts as an XON to allow or re-enable unit to continue
sending characters:
HOST SEND
HOST SEND
AZS<cr> Non-network
AZ[yyyyy]S<cr> Network
RESPONSE - none